  • NEVER COOL A HANDPIECE WITH WATER.
  • LET THE HANDPIECE COOL TO ROOM TEMPERATURE FOR A MINIMUM OF 30 MINUTES BEFORE USE.

Handpiece Maintenance:

High Speed Maintenance Dental Handpiece Repair and Maintenance Service

  1. Disassemble handpiece from tubing.
  2. Clean exterior surfaces and fiber-optics with alcohol. (Do not use harsh cleaners or ultrasonic cleaner.)
  3. Insert a bur, and spray “The Dentist’s Choice” lubricant into air drive port.
  4. Rotate bur between the thumb and forefinger.
  5. Re-Attach handpiece to tubing, cover with paper towel and run.
  6. Examine discharged oil on towel. If discolored, repeat lubrication procedure until discharge becomes clear.
  7. Clean off excess surface oil.
  8. Run handpiece for 20 to 30 seconds.
  9. Repeat, in 20 – 30 second bursts, until all excess oil is expelled from inside handpiece.
  10. Autoclave handpiece following manufacturer’s recommended procedures.
  11. Remove handpiece and let cool. (Handpiece must be at room temperature before proceeding.)
  12. Insert a bur, and spray lubricant into air drive port.
  13. Rotate bur between the thumb and forefinger.
  14. Attach handpiece to tubing, and run for 20 to 30 seconds.

High Speed Lubrication.

Dental Tool Lubricants by Dentist's Choice

Proper Maintenance Saves time, money and wear and tear of tools.

  1. Lubricate with Once-A-Day lubricant before autoclaving.
  2. Run handpiece 30 seconds to clear excess oil.
  3. Autoclave handpiece following manufacturer’s recommended procedures.
  4. Remove handpiece and let cool. (Handpiece must be at room temperature before proceeding.)
  5. Lubricate as noted above.
  6. Spin off excess oil.

High Speed Tips

  1. High speed handpieces cannot be over-lubricated.
  2. Keep manufacturer’s recommended procedures on file for reference.
  3. Never use pin oil or syringe oil on high speed handpieces.
  4. Lubricate before autoclaving.
  5. Set air pressure to Manufacturer’s specifications.

Lube Free High Speed Maintenance

  1. Lubrication is not required.
  2. To clean, remove bur and blow air into chuck.
  3. Insert bur. (Never operate without a bur.)
  4. Run for 20 seconds to flush air and water lines.
  5. Clean exterior surface with alcohol.
  6. Dry thoroughly.
  7. Autoclave handpiece following recommended procedures. (Never submerge handpiece in chemical disinfectants, ultrasonic cleaner or cold sterilants.)
  8. Allow 30 minutes to cool to room temperature.

Low Speed Maintenance

  1. Clean exterior surface with alcohol. (Do not use harsh cleaners, ultrasonic cleaners, or introduce cleaner into motor.)
  2. Insert a bur and tighten.
  3. Remove handpiece from tubing.
  4. Add 2 to 3 drops of oil into drive air tube.
  5. Attach handpiece to tubing, and run at high speed for 10 to 20 seconds.
  6. Consult with manufacturer before autoclaving any low speed handpieces for recommended procedures.

Low Speed Lubrication

  1. Add 2 to 3 drops of oil into drive tube.
  2. Insert a bur.
  3. Run handpiece 10 to 20 seconds.

Low Speed Tips

  1. Lubricate after each use.
  2. Keep manufacturer’s recommended procedures on file for reference.
  3. Set air pressure to manufacturer’s specifications.
  4. Use high viscosity lubricating and cleaning oil.